Miele PG 8583 CD Operating Instructions Manual page 113

Laboratory glassware washer
Hide thumbs Also See for PG 8583 CD:
Table of Contents

Advertisement

Setting the
temperature and
time
Setting
temperature 1
Setting drying
time 1
Setting
temperature 2
The drying phase consists of two blocks. The temperature and
duration (holding time) must be set for each block.
The first block (temperature 1 and drying time 1) is not assigned in all
programs, but if necessary can be set up by Miele Service.
   ...
Temperature 1
         
The temperature is set in 5° increments. The possible range is shown
in the bottom line of the display.
Use the  (higher) and  (lower) arrow buttons to set the
temperature.
Press OK to save the setting.
   ...
Drying time 1
         
The holding time is set in 1 minute increments. The possible range is
shown in the bottom line of the display.
Use the  (higher) and  (lower) arrow buttons to set the holding
time.
Press OK to save the setting.
   ...
Temperature 2
         
The temperature is set in 5° increments. The possible range is shown
in the bottom line of the display.
Use the  (higher) and  (lower) arrow buttons to set the
temperature.
Press OK to save the setting.
Temperature 1
75
°C
(50 - 115
°C)
Drying time 1
30
Min
(0 - 120
Min)
Temperature 2
75
°C
(50 - 115
°C)
Program settings
113