Miele PG 8583 CD Operating Instructions Manual page 15

Laboratory glassware washer
Hide thumbs Also See for PG 8583 CD:
Table of Contents

Advertisement

A damaged or leaking machine is dangerous and poses a safety
hazard. Immediately disconnect the machine at the power switch and
contact the Miele Service Department.
Label machines which have been taken out of operation and lock
them to prevent them being switched on again without authorization.
The machine may only be put back into operation once it has been
successfully repaired by Miele Service or by an appropriately qualified
specialist.
Personnel operating the machine should be trained on a regular
basis. Untrained personnel must not be allowed access to the
machine or its controls.
Only use process chemicals which have been approved by their
manufacturer for the relevant application. The manufacturer of the
process chemicals is liable for any negative influences on the material
of the load items and the machine.
Always exercise caution when handling the process chemicals for
this machine. These products may contain irritant, corrosive or toxic
ingredients.
Always comply with safety requirements and the manufacturer's
safety instructions (see safety data sheets)!
Use protective eyewear and gloves!
The machine is designed to operate with water and the
recommended process chemicals only. Organic solvents or
flammable liquid agents must not be used in it!
This could cause an explosion, property damage due to the
destruction of rubber and plastic components, and the resulting
leakage of liquids.
The water in the cabinet must not be used as drinking water.
Do not lift the machine by protruding parts such as the control
panel or the opened service flap as these could be damaged or torn
off.
Do not sit or lean on the opened door. This could cause the
machine to tip or become damaged.
Be careful when sorting load items with sharp, pointed ends.
Position them in the machine so that you will not hurt yourself or
create a danger for others.
Broken glass can result in serious injury when loading or
unloading. Broken glass items must not be processed in the machine.
When operating the machine, beware of the high temperatures
involved. If you bypass the electrical lock to open the door, there is a
danger of scalding and heat or chemical burns. If disinfectants have
been used, there is also the danger of inhaling toxic vapour.