Program settings
Repeating if limit
value is exceeded
Calibrating the
conductivity
meter
106
If the conductivity exceeds the limit value for the water intake or the
water drain, the measurement can be repeated.
If the limit value of the water intake is too high, the water is pumped
out and the water intake is repeated. if the conductivity at the end of
the rinse cycle is above the limit value for the drained water, the entire
final rinse is repeated.
If the value exceeds the limit again, the program will be cancelled with
a error message. The message will be displayed both on the display
and the cycle report.
   ...
Conductivity limit
         
Water intake
             
                 
Number of repeats
Adjustment is in increments of 1. The possible range is shown in the
bottom line.
Use the arrow buttons  (higher) and  (lower) to set the number of
repetitions.
Press OK to save the setting.
The conductivity meter must be recalibrated by Miele Service at
regular intervals, e.g. during servicing. Sometimes calibration may
also be necessary outside the service cycle.
If this message appears, contact the Miele Service Department.
Water drainage
or
Number of repeats
1
(0 - 1)
Conductivity module
requires calibration