Replacing The Hepa Filter - Miele PG 8583 CD Operating Instructions Manual

Laboratory glassware washer
Hide thumbs Also See for PG 8583 CD:
Table of Contents

Advertisement

Maintenance
Replacing the
HEPA filter
128
Replace with an Original Miele HEPA Filter classification 14 for
optimum performance.
If possible, Miele Service should replace the HEPA filter as part of a
maintenance visit. If this is not possible, you can change the filter as
follows:
Open the drawer in the side unit and remove the filter grille and the
coarse filter.
Detach the coarse filter housing by unscrewing the retaining screws
and pushing them upward.
Remove the coarse filter housing.
Remove the HEPA filter from its holder and insert a new one.
Reinsert the coarse filter housing and tighten the retaining screws.
Reinsert the coarse filter and the filter grille and close the drawer in
the side unit.
Whenever the HEPA filter is replaced, the operating hours counter
must be reset (see "Resetting the operating hours counter").

Advertisement

Table of Contents
loading

Table of Contents