Performance Check - Miele PG 8583 CD Operating Instructions Manual

Washer-disinfector for laboratory glassware and laboratory utensils
Hide thumbs Also See for PG 8583 CD:
Table of Contents

Advertisement

Test point for
measuring
sensors
 The operating hours counter must only be reset when the filter
has been changed.
Reset
Yes
No
Yes
-
The operating hours counter will be reset for the new filter.
No
-
The counter will not be reset.
Select an option using the  and  arrow buttons.
Confirm your selection with OK.

Performance check

Adequate processing performance must be regularly confirmed by
the user.
The sensor test point for validation is located at the front right on the
top of the machine under the lid or the worktop. To reach the access
point, the lid of the machine must be removed or the machine must
be pulled out from under the worktop.
Open the door.
Unscrew the fixing screws.
Then remove the safety screws on the back of the machine from
the lid and lift the lid to remove it.
Or
Pull the washer-disinfector out by approx. 15 cm from under the
worktop.
Maintenance
131