Changing the
HEPA filter
Replace with an Original Miele HEPA filter classification 14 for
optimum performance.
If possible, the Miele Customer Service Department should replace
the HEPA filter as part of a maintenance visit. If this is not possible,
you can change the filter as follows:
Open the drawer in the side unit and remove the filter grille and the
coarse filter.
Detach the coarse filter housing by unscrewing the fixing screws
and pushing them upwards.
Remove the coarse filter housing.
Remove the HEPA filter from its holder and insert a new one.
Reinsert the coarse filter housing and tighten the fixing screws.
Reinsert the coarse filter and the filter grille and close the drawer in
the side unit.
Whenever the HEPA filter is replaced, the operating hours counter
must be reset (see "Resetting the operating hours counter").