Changing the HEPA filter

The air filter for the internal drying unit has a limited lifespan and has
to be replaced at regular intervals, e.g. when the following message
appears:
Fit a new HEPA filter!
Replace with an Original Miele HEPA filter classification 13 for
optimum performance.
Open the service panel in the plinth.
Push the retaining clips outwards to release the HEPA filter.
Grasp the indentations on the sides and pull the filter forwards.
Insert a new HEPA filter, making sure that it locates securely in the
retaining clips.
Close the service panel.
Whenever the HEPA filter is replaced the operating hours counter
must be reset.