2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit
2.4 Preparation of the dentist unit
2.4.1 Preparation of the handle and holder pad/ anti-
slip mat
Cleaning and disinfection of the handle and holder pad/
antislip mat (T table / S table)
▶ Press the push button and pull off the handle.
▶ Clean the handle under flowing water.
▶ Disinfect the white and grey handle by wiping or thermodisinfection (minor
discolouration may occur).
▶ Sterilise the holder pad and anti-slip mat by thermodisinfection.
Sterilisation of the handle and holder pad/ anti-slip mat
(T table / S table)
Damage caused by sterilisation.
▶ Do not sterilise the white standard handle.
▶ Sterilise the holder pad/ anti-slip mat.
The grey handle (Mat. no. 1.002.4489) can be sterilised (see symbol on the
handle).
Care instructions ESTETICA E70 Vision / E80 Vision
CAUTION
11 / 48