6 Preparation methods DIN EN ISO 17664

CAUTION
|

6.1 Cleaning

6 Preparation methods DIN EN ISO 17664
6.1 Cleaning

6.1.1 Clean unit

The wide variety of drugs and chemicals used in dental practices means that da‐
mage may occur to painted surfaces or plastics if anything is allowed to drip onto
them.
Tests have shown that surfaces cannot be protected entirely against all substances
available on the market.
As damage to surfaces is very much dependent upon the reaction times of these
substances, it is essential for any spilled substances to be wiped away immediately
using a damp cloth.
The following cleansers may not be used:
▪ Strongly alkaline washing solutions
▪ Acids
▪ Ammonia-containing cleaning agents
▪ Abrasives of any kind
▪ Solvents on plastic parts
Damage from liquids.
Malfunctions of electrical components.
▶ Cover openings in the product to prevent the penetration of liquids.
▶ Remove liquids from the interior of the device.
▶ Switch off the instrument's main switch.
▶ Clean surface with a soft cloth and a mild cleaning fluid.
Note
If there is any dirt or contamination on the unit that you are unable to remove, the
unit should be inspected and cleaned by a KaVo Service Technician or KaVo-trai‐
ned service technician.