The microprocessor-controlled hydraulic unit offers the following advantages
Approximation of the physiological gait pattern
Stability while standing and walking
Adaptation of product characteristics to various surfaces, inclines, gait situations and walking speeds

2.3 Combination possibilities

This product can be combined with the following Ottobock components:
Adapters
4R104=60 double adapter, sliding
4R104=75 double adapter, sliding
Rotation adapter: 4R57, 4R57=*
4R41 lamination anchor with pyramid receiver
4R43 lamination anchor with threaded connector
4R89 lamination anchor with pyramid adapter
4R111=N lamination anchor with threaded connect­
or
Tube adapter
AXON tube adapter: 2R20
AXON tube adapter with torsion unit: 2R21
Cosmetic cover/protector
Foam cover: 3S26
4X880=* Genium Protective Cover
Prosthetic feet
The maximum allowable patient weight depends on the foot size.
1M10 Adjust
1A30 Greissinger plus
1C30 Trias
1D35 Dynamic Motion
1C40 C-Walk
1C60 Triton
1C61 Triton Vertical Shock
1C62 Triton Harmony
1C63 Triton Low Profile
Depending on the patient's weight, this knee joint may be combined with the 1C63 Triton Low Profile pros­
thetic foot and the 1A1-1 Empower prosthetic foot in the following foot sizes only (see following table).
Please contact Ottobock customer service if you would like a combination outside the approved ranges.
Body weight
up to 116 kg (up to 256 lbs)
117 kg to 125 kg (257 lbs to 275 lbs)
126 kg to 130 kg (276 lbs to 286 lbs)
131 kg to 150 kg (287 lbs to 330 lbs)

2.3.1 Combination with an osseointegrated implant system

This product can be connected to a socket or to an osseointegrated, percutaneous implant system.
In case of connection to an implant system, verify that the manufacturer of the implant system and the manufactur­
ers of the corresponding exoprosthetic components/adapters also permit this combination. It must be ensured that
all indications/contraindications, the field of application, the conditions of use and all safety instructions are com­
plied with for the implant system, corresponding exoprosthetic components, corresponding adapters and for the
knee joint.
Among other things, this relates to the body weight, mobility grade, type of activity, load capacity of the implant and
bone anchoring, freedom from pain under functional load and compliance with the permissible ambient conditions
(see Page 96).
Genium 3B1-3/3B1-3=ST
4R111 lamination anchor with pyramid receiver
4R116 lamination anchor with pyramid adapter
Lamination anchor with pyramid receiver and
angled arm: 4R119
4R40 torsion adapter
4R118 adapter plate
Quickchange: 4R10
1C64 Triton Heavy Duty
1E56 Axtion
1E57 Lo Rider
Challenger: 1E95
1B1 Meridium
Meridium: 1B1-2
1C66 Triton smart ankle
1C10 Terion
Empower: 1A1-1
Foot sizes 1C63 [cm]
21 to 30
21 to 30
21 to 28
21 to 28
Product description
Foot sizes 1A1-1 [cm]
25 to 30
25 to 27
25 to 27