Reset User Calibration Defaults; Calibration Menu - Thermo Scientific 410i Instruction Manual

Table of Contents

Advertisement

Operation

Calibration Menu

Reset User
Calibration Defaults
Calibration
3-20
Model 410i Instruction Manual
In the Main Menu, choose Calibration Factors > O
The Reset User Calibration Defaults screen allows the user to reset the
calibration configuration values to factory defaults.
In the Main Menu, Calibration Factors > Reset User Cal Defaults.
RESTORE DEFAULT CAL:
RANGE
AVG
DIAGS ALARM
The Calibration menu is used to calibrate zero and span. The Calibrate O
Background menu item and the Calibrate O
Menu
displayed only when the internal O
calibration menu is similar for the single, dual, and auto range mode as
shown below. The only difference between the screens are the words "HI"
and "LO" to indicate which range is displayed. The dual and auto range
modes have two CO
be calibrated separately. This is necessary if the two ranges used are not
close to one another. For example, a low CO
CO
range of 1000 ppm. For more information about calibration, see
2
Chapter 4, "Calibration".
In the Main Menu, choose Calibration.
O2 COEFFICIENT:
O2:
SET COEF TO:

INC/DEC
SAVE VALUE
RANGE
AVG
DIAGS ALARM
RESTORE
span factors (high and low). This allows each range to
2
20.80
1.00
?
RESTORE DEFAULT CAL:
ARE YOU SURE YOU WANT
PRESS
TO CONFIRM RESTORE
RANGE
AVG
DIAGS ALARM
Coefficient menu item are
2
sensor option is installed.The
2
range of 200 ppm and a high
2
Coefficient.
2
RESTORE
TO?x
2