Temperature Calibration; Analog Output Calibration - Thermo Scientific 410i Instruction Manual

Table of Contents

Advertisement

Temperature
Calibration
Analog Output
Calibration
Thermo Fisher Scientific
In the Main Menu, choose Service > Preamp Calibration.
PREAMP BOARD CAL:
SAMPLE:
REFERENCE:
SET TO:

CHANGE VALUE
RANGE
AVG
The Temperature Calibration screen allows the user to view and set the
ambient temperature sensor offset, which basically adjusts the temperature
sensor so that it reads correctly. The temperature calibration is visible only
when the instrument is in service mode. For more information on the
service mode, see
"Service
Note This adjustment should only be performed by an instrument service
technician.
Note Wait at least 30 seconds for the reading to stabilize before saving the
value.
In the Main Menu, choose Service > Temperature Calibration.
CALIBRATE AMBIENT TEMP:
CURRENTLY:
SET TO:

CHANGE VALUE
RANGE
AVG
The Analog Output Calibration menu provides access to the 6 voltage
channels and 6 current channels for calibration. Current channels are
visible only if the I/O expansion board is installed. The analog output
calibration menu is visible only when the instrument is in service mode.
For more information on the service mode, see
the chapter.
Note This adjustment should only be performed by an instrument service
technician.
127807 Hz
148256 Hz
128
SAVE
DIAGS ALARM
Mode" earlier in the chapter.
o
25.0
C
o
037.5
C ?

MOVE CURSOR
SAVE
DIAGS ALARM
Operation
Service Menu
?
"Service
Mode" earlier in
Model 410i Instruction Manual
3-85