alarm motor speed min
alarm motor speed max
These commands report the motor speed alarm minimum and maximum
value current settings. The following example reports that the motor speed
alarm minimum value is 99.8 %.
Send:
alarm motor speed min
Receive:
alarm motor speed min 99.8
set alarm motor speed min value
set alarm motor speed max value
These commands set the motor speed alarm minimum and maximum
values to value, where value is a floating-point number representing motor
speed alarm limits in minutes. The following example sets the motor speed
alarm maximum value to 100.2 %.
Send:
set alarm motor speed max 100.2
Receive:
set alarm motor speed max 100.2 ok
alarm pressure min
alarm pressure max
These commands report the pressure alarm minimum and maximum value
current settings. The following example reports that the pressure alarm
minimum value is 650 mmHg.
Send:
alarm pressure min
Receive:
alarm pressure min 650 mmHg
set alarm pressure min value
set alarm pressure max value
These commands set the pressure alarm minimum and maximum values to
value, where value is a floating-point number representing pressure alarm
limits in millimeters of mercury. The following example sets the pressure
alarm maximum value to 790 mmHg.
Send:
set alarm pressure max 790
Receive:
set alarm pressure max 790 ok
alarm sample flow min
alarm sample flow max
These commands report the sample flow alarm minimum and maximum
value current settings. The following example reports that the sample flow
alarm minimum value is 0.350 L/min.
Send:
alarm sample flow min
Receive:
alarm sample flow min 0.350 l/min
C-Link Protocol Commands
Model 410i Instruction Manual
Alarms
B-17