Thermo Scientific 410i Instruction Manual page 190

Table of Contents

Advertisement

Servicing
Optical Switch Replacement
Optical Switch
Replacement
7-18
Model 410i Instruction Manual
Use the following procedure to replace the optical switch. Refer to
7–11
or
Figure 7–12
as appropriate.
Equipment Required:
Optical switch
Flatblade screwdriver, 1/4-inch and 3/16-inch
Philips screwdriver
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
1. Turn the instrument OFF, unplug the power cord, and remove the
cover.
2. Remove the optical bench following the
procedure in this chapter.
3. Turn the optical bench over, remove the two screws securing the
optical switch assembly to the case, and remove the optical switch.
4. Install the new optical switch by following the previous steps in reverse.
5. Calibrate the instrument. Refer to the "Calibration" chapter in this
manual.
Figure
"Optical Bench
Replacement"
Thermo Fisher Scientific