Dilution Ratio

Setup Run Screens

Thermo Fisher Scientific
O2 CORRECTION CONC:
CURRENTLY:
SET TO:

CHANGE VALUE
RANGE
AVG
The Dilution Ratio screen allows the user to view and set the dilution ratio.
Acceptable values are 1–500: 1. The default is 1:1. When this value is set,
the dilution ratio is applied to all concentration measurements. This screen
is only accessible if the dilution ratio option is installed and when the
instrument is in service mode. For more information on the service mode,
see
"Service
Mode" earlier in the chapter.
In the Main Menu, choose Service > Dilution Ratio.
DILUTION RATIO:
CURRENTLY:
SET TO:

CHANGE VALUE
RANGE
AVG
The Setup Run Screens menu is used to define and edit up to five custom
Run screens. When the instrument is turned on, screen 1 (or zero) is the
default screen. The Setup Run Screens menu is visible only when the
instrument is in the Service mode. For more information on the service
mode, see
"Service
Mode" earlier in the chapter.
In the Main Menu, choose Service > Setup Run Screens.