Figure 7-4. Removing The Measurement Case Assembly And Lowering The Partition Panel (410I Standard Shown) - Thermo Scientific 410i Instruction Manual

Table of Contents

Advertisement

Removing the
Measurement Case
Assembly and Lowering
the Partition Panel
REMOVE SCREWS
Loosen
Partition Panel
Captive
Screw
Remove Screw
Thermo Fisher Scientific
The measurement case assembly can be removed and the partition panel
can be lowered to improve access to connectors and components. Refer to
the following steps when a procedure requires lowering the partition panel
(see
Figure
Pass Through
Connectors
Remove Screws
Figure 7–4. Removing the Measurement Case Assembly and Lowering the
Partition Panel (410i Standard shown)
Equipment Required:
Philips screwdriver
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
1. Turn the instrument OFF and unplug the power cord.
2. If the instrument is mounted in a rack, remove it from the rack.
Removing the Measurement Case Assembly and Lowering the Partition Panel
7–4).
Remove Screw
Servicing
SLIDE CASE ASSEMBLY OUT
LOWER PANEL
Remove Hidden Screw
Model 410i Instruction Manual
7-9