C-Link Protocol Commands
Datalogging
B-22
Model 410i Instruction Manual
Send:
erec
Receive:
erec
09:48 04-06-05 flags 9C040510 CO2 0.000 4 loCO2 -0.002
4 s/r 0.000 los/r 0.902 biasv -112.668 agci 96.500 intt
34.023 cht 47.995 smpfl 0.000 pres 0.000 avgt1 10 avgt2
10 CO2bkg -0.000 CO2coef 1.000 loCO2coef 1.000 CO2range
10000000.000 loCO2range 10000000.000 motor 100.000
lrec
srec
lrec rec num
srec rec num
lrxy rec num
srxy rec num
lrec aa:bb oo-pp-qq yy
srec aa:bb oo-pp-qq yy
rec = The starting record index number (1=most recent)
num = The number of records to return (1 to 10)
x = | 0 | 1 |
: Reply termination format (see "set format format"
command)
y = | 0 | 1 | 2 | : Output format (see "set lrec/srec format format"
command)
aa = hours (01 to 23)
bb = minutes (01 to 59)
oo = month (01 to 12)
pp = day (01 to 31)
qq = year
These commands output lrec or srec logged data. The output format is
determined by the "set lrec format", and "set srec format" commands. The
logging time is determined by the "set lrec per" and "set srec per"
commands.
In the following example, assume there are 740 lrecs currently stored in
memory. When the command "lrec 100 5" is sent, the instrument counts
back 100 records from the last record collected (record 740), and then
returns 5 records: 640, 641, 642, 643, and 644. For details on how to
decode the flag fields within these records, see
Send:
lrec 5
Receive:
lrec 100 5
10:15 05-12-03 flags 9c040000 CO2 7349E+0 loCO2 5994E+0
intt 33.2 cht 44.7 pres 758.9 smplfl 1.085 speed 100.0
biasv -115.5 intensity 1999940
10:15 05-12-03 flags 9c040000 CO2 7349E+0 loCO2 5994E+0
intt 33.2 cht 44.7 pres 758.9 smplfl 1.085 speed 100.0
biasv -115.5 intensity 1999940
Figure