Figure 7-15. Replacing The Pump - Thermo Scientific 410i Instruction Manual

Table of Contents

Advertisement

Servicing
Pressure Transducer Replacement
Pump Bracket
Pump Mounting Screws (2)
(Bottom)
Pressure Transducer
Replacement
7-24
Model 410i Instruction Manual
3. Remove both lines from the pump.
4. Loosen the four captive screws holding the pump bracket to the shock
mounts and remove the pump assembly and the shock mounts.
5. Invert the pump assembly, remove the two pump mounting screws
located on the bottom side of the pump bracket, and remove the
bracket from the pump.
6. Install the new pump by following the previous steps in reverse.
7. Perform a leak test as described in the
chapter.
Pump
Figure 7–15. Replacing the Pump
Use the following procedure to replace the pressure transducer
16).
Equipment Required:
Pressure transducer
Philips screwdriver, #2
"Preventive
Maintenance"
Captive Screws (4)
Shock Mounts (4))
Floor Plate
(Figure 7–
Thermo Fisher Scientific