Diagnostics
Thermo Fisher Scientific
set alarm trig conc o2 value
This command sets the O
value is set to either floor or ceiling, according to
example sets the O
concentration minimum alarm trigger to ceiling.
2
Send:
set alarm trig conc o2 1
Receive:
set alarm trig conc o2 1 ok
agc int
This command reports the current intensity of the reference channel AGC
circuit. The following example reports that the current AGC intensity is
200,000 Hz.
Send:
agc int
Receive:
agc int 200000 Hz
bias voltage
This command reports the current IR bias supply voltage. The following
example reports that the bias voltage is -102.3 volts
Send:
bias voltage
Receive:
bias voltage -102.3 V
diag volt mb
This command reports the diagnostic voltage measurements on the
motherboard. The sequence of voltages is: Positive 24, positive 15, positive
5, positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag volt mb
Receive:
diag volt mb 24.1 14.9 4.9 3.2 -3.2
diag volt mib
This command reports the diagnostic voltage measurements on the
measurement interface board. The sequence of voltages is: Positive 24,
positive 15, negative 15, positive 5, positive 3.3, positive 18 IR, positive 18
MOT, and VBIAS. Each voltage value is separated by a space.
Send:
diag volt mib
Receive:
diag volt mib 24.1 14.9 -14.9 4.9 3.2 17.9 17.9
diag volt iob
This command reports the diagnostic voltage measurements on the I/O
expansion board. The sequence of voltages is: Positive 24, positive 5,
positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag volt iob
Receive:
diag volt iob 24.1 4.9 3.2 -3.2
concentration alarm minimum value, where
2
Model 410i Instruction Manual
C-Link Protocol Commands
Diagnostics
Table
B–4. The following
B-19