Multi-Point Calibration; Calibrate Point 1/2/3 - Thermo Scientific 410i Instruction Manual

Table of Contents

Advertisement

Multi-Point
Calibration

Calibrate Point 1/2/3

Thermo Fisher Scientific
INITIAL S/R RATIO:
CURRENTLY:
MEASURED:
SET TO:

CHANGE VALUE
RANGE
AVG
Up to three gas concentrations (cal-points) for each range may be calibrated
to using the following steps. Three cal-points will give the most accurate
readings over the entire range. The calibration process is sequential and will
work properly if all steps are followed in order. The following example
shows the multi-point calibration screen in single range mode. In dual or
auto range modes, "HI" or "LO" is displayed to indicate the calibration of
the high or low concentrations and function the same way.
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Multipoint Calibration.
MULTI-POINT CAL:
>CALIBRATE POINT 1
CALIBRATE POINT 2
CALIBRATE POINT 3
COEFFICIENTS
DEFAULT COEFFICIENTS
RANGE
AVG
The Calibrate Point 1 screen allows the user to view and set the selected
calibration point. The calibrate 2 and calibrate 3 screens function the same
way.
In the Main Menu, choose Service > Multipoint Cal > Calibrate 1, 2
or 3.
1.079620
0.930966
0.931524

MOVE CURSOR
SAVE
DIAGS ALARM
DIAGS ALARM
Operation
Service Menu
?
Model 410i Instruction Manual
3-83