4.1 Conditions on site

T10FS torque flanges are protected to IP54 according to EN 60529. They
must be protected against coarse dirt particles, dust, oil, solvents and
humidity. During operation, the prevailing safety regulations for the security of
personnel must be observed (see "Safety instructions").
There is wide ranging compensation for the effects of temperature on the
output and zero signals of the T10FS torque flange (see Specifications on
Page 67). This compensation is carried out at static temperatures in extensive
furnace processes. This guarantees that the circumstances can be
reproduced and the properties of the transducer can be reconstructed at any
time.
If there are no static temperature ratios, for example, because of the
temperature differences between the measuring body and the flange, the
values given in the specifications can be exceeded. So for accurate
measurements, static temperature conditions must then be obtained by
cooling or heating depending on the application. As an alternative, check
thermal decoupling by means of heat radiating elements such as multi‐disk
couplings.

4.2 Installation orientation

The measurement flange can be mounted in any position. With clockwise
torque, the output frequency is 10 kHz to 15 kHz. With HBM amplifiers or with
the "voltage output" option, a positive output signal (0 V ... +10 V) is present.
In the case of the speed measuring system, an arrow is attached to the head
of the sensor to clearly define the direction of rotation. If the measurement
flange moves in the direction of the arrow, connected HBM measuring
amplifiers deliver a positive output signal (0 V ... +10 V).
HBM
T10FS
A0785-1.4 en