Conditions At The Site Of Installation; Mounting Position - HBM T10F Mounting Instructions

Torque flange
Hide thumbs Also See for T10F:
Table of Contents

Advertisement

10
3.1

Conditions at the site of installation

The T10F Torque Flanges are protected to IP54 according to EN 60 529. The
measuring hubs must be protected from dirt, dust, oil, solvants and humidity.
During operation, the prevailing safety regulations for the security of
personnel must be observed (see "Notes on safety").
The output signal and zero point of the T10F Torque Measurement Flange are
compensated for the effects of temperature variations between wide limits
(see Technical Data on page 36). This compensation is carried out at static
temperatures. This ensures that reproducible relationships are present and
that the characteristics of the transducers can be duplicated at any time. If
static temperature relationships do not prevail, e.g. due to temperature
differences between the measuring element and the adapter flange, the
values specified in the technical data can be exceeded. Then for accurate
measurements, static temperature relationships must be obtained by cooling
or or heating depending on the application. Alternatively, temperature
decoupling must be checked, e.g. by heat radiating elements such as
laminated couplings.
3.2

Mounting position

The torque flange can be mounted in any position. With clockwise torque, the
output frequency is 10...15kHz. With HBM amplifiers or with the "Voltage
output" option, a positive output signal (0...+10V) is present.