Agilent Technologies 8453 User Manual, Installing And Operating Manual page 95

Dissolution testing system
Hide thumbs Also See for 8453:
Table of Contents

Advertisement

4 Slide a ferrule over the end of the tubing—the ferrules are beveled, make
Figure 37
5 Pull the tubing through fitting, o-ring and ferrule so that it protrudes about
Figure 38
6 Connect the tubing with the loose fitting, o-ring and ferrule to a tubing with
Figure 39
Always hold the nut extender tool at the end with the finger grip—optimum tightness is
N O T E
reached when the nut extender tool begins to slip.
Agilent 8453 Dissolution Testing System Installing and Operating Manual
sure the narrow end of the bevel points towards the o-ring, see
Orientation of Fitting, O-ring and Ferrule on Tubing
4 mm (3/16 inch) beyond the ferrule, see
Fitting Ready for Connection to Other Tubing
a fixed fitting and use the nut extender tool to tighten the fittings, see
Figure
39.
Tightening the Fittings
Using the Valve Unit and Valve-pump Controller
Troubleshooting Your Valve System
Figure
4 mm
Figure
38.
4
37.
95