Using the Valve Unit and Valve-pump Controller
Troubleshooting Your Valve System
Changing Valve
Change the complete valve assembly when the following symptoms still occur
after cleaning the valve (as described in
page 91) and troubleshooting the tubing (as described in
Connectors at Valve"
page 90):
• valve leaking;
• air bubbles in tubing leading from valve;
• >0.5% cross contamination between samples;
• valve has stuck (see
To remove the Valve
1 Bring the valve to position 1 if it operates.
2 Turn off the valve-pump controller.
3 Remove the two screws holding the valve to the valve drive assembly, at the
4 Remove the valve. You may have to move the valve left, right, up or down to
Valve drive assembly
Figure 35
92
on page 89 and
front.
release it from the shaft.
Removing the Valve
Agilent 8453 Dissolution Testing System Installing and Operating Manual
"Cleaning Tubing and Valve"
"Contamination Between Ports"
"Valve Unit does not Operate"
"Leaking Tubing
on page 88).
Valve
Screws
on