Operating Your Pump

You may operate the pump remotely or manually with or without the
peristaltic pump remote cable. Before using the pump with the remote cable
adjust the flow rate as described in

Adjusting Flow Rate

1 Remove the peristaltic pump remote cable from the remote connector at the
2 Turn on the line power at the line switch on the front of the pump.
3 Place the left end of the tubing in a reservoir of liquid.
4 Place the right end of the tubing in a empty waste beaker.
5 Press the control switch on the front panel to the right to pump clockwise.
6 Set the flow control to about 40.
7 Change the flow setting for the desired flow rate according to
Table 19
Flow Setting
40
60
80
8 Reconnect the peristaltic pump remote cable to the remote connector to
For operating the pump through the peristaltic pump remote cable from the
Agilent 8453 spectrophotometer, see the appropriate manual.
Agilent 8453 Dissolution Testing System Installing and Operating Manual
rear of the pump.
Flow Settings
Approximate Flow Rate (Water)
6.0 ml/min
8.5 ml/min
11.5 ml/min
enable remote operation from the Agilent 8453 spectrophotometer.
Using the Peristaltic Pumps 1VS and 8VS
Operating Your Pump
"Adjusting Flow Rate"
3
on page 73.
Table
19.