3 Connect the dissolution probes to the flow cells—connect the 2.0 m tubing,
Colored heat-shrunk
sheath on tubing
Long cell fitting
Union
Dissolution probe
Figure 7
4 Connect the 2.0 m tubing with the colored heat-shrunk sheath at one end to
5 Connect the union to a conical adapter.
6 Insert a conical adapter into the pump tubing on the outlet side of the
7 Insert the open ends of the tubing into the vessels.
8 Bundle your tubing together with the spiral tubing wrap to keep it from
Agilent 8453 Dissolution Testing System Installing and Operating Manual
see
Figure
7, with the long cell fitting to the inlet (marked with
flow cell. Then, connect the fitting to the union attached to the dissolution
probe.
Connecting Dissolution Probes to Fitting
a union.
pump.
kinking.
Installing a Multicell-transport-based Sampling System
Setting Up
2
) of the
33