Using the Peristaltic Pumps 1VS and 8VS
Maintaining Your Pump
3 Remove gear box mounting-plate—see
Removing Drive Belt
1 Using a slot screwdriver, push the drive belt off the toothed-drive wheel
2 Remove the drive belt from the toothed-wheel of the gear box.
Figure 33
78
of the screws holes to aid in locating the screws. Using a Phillips
screwdriver, unscrew the four screws holding the gear box mounting-plate
to the front of the pump. This releases the gear box and drive belt from the
front panel.
(Figure
33), at the same time turning the toothed-drive wheel of the gear
box
(Figure
33).
Drive Belt of the Pump
Agilent 8453 Dissolution Testing System Installing and Operating Manual
Figure 31
on page 76 for the position