Line Fuse
The line fuse is located in the line filter, accessible from the rear of the
valve-pump controller.
Fuses F1, F2
These fuses are located inside the valve-pump controller on the rear panel
connector board, see
Figure 41
Agilent 8453 Dissolution Testing System Installing and Operating Manual
T630 mA (2110-0450) is used for 115 V operation.
T315 mA (2110-0449) is used for 230 V operation.
Figure
F1: T2000 mA (2110-0601) for the 5 V DC line.
F2: T1000 mA (2110-0457) for the 24 V AC line.
F1
F2
Internal Fuses
Using the Valve Unit and Valve-pump Controller
Servicing Your Valve System
41.
4