Installing Your Eight-Position Multicell Transport; Installing Your Peristaltic Pump 8Vs - Agilent Technologies 8453 User Manual, Installing And Operating Manual

Dissolution testing system
Hide thumbs Also See for 8453:
Table of Contents

Advertisement

2
Setting Up
Installing a Multicell-transport-based Sampling System
Position 1 is at the left side of the multicell transport.
N O T E
Position 8 is at the right side of the multicell transport.

Installing Your Eight-position Multicell Transport

To install your eight-position multicell transport in the Agilent 8453
spectrophotometer, refer to your Multicell Transport for Agilent 8453
UV-visible Spectrophotometer Operator's Manual.

Installing Your Peristaltic Pump 8VS

Each peristaltic pump 8VS is identified by a serial number on a label at the
rear. Write down the serial number of your pump here:
_____________________________________________________________________
Installing the pump comprises three stages:
Stage 1: Installing the Cables
The voltage selector and cable connectors are at the rear panel.
Make sure line-power cord is disconnected before changing line-voltage setting.
WA R N I N G
1 Set the correct voltage option (115 V or 230 V) for operation from your line
2 Check the correct fuse is installed.
3 Push connector of power cord into line-power socket.
28
"Stage 1: Installing the Cables"
"Stage 2: Installing the Pump Tubing"
"Stage 3: Installing the Tubing Cassette"
power supply.
Agilent 8453 Dissolution Testing System Installing and Operating Manual
on page 28,
on page 29, and
on page 30.