Setting Up
Setting Up Your Sampling System

Dissolution Bath

A dissolution probe kit and a packet of 40 filter tips are part of the valve and
multicell transport sampling systems.
Union
Fitting
Probe
Filter
Figure 1
A dissolution probe kit consists of eight fittings with a probe (25 cm, 1/8-inch
od) and union attached.
1 Cut the probes to a suitable length for your dissolution bath, if necessary.
2 Attach a filter to each probe.
If you are using other filters with smaller pore size than 45 µm they may reduce the flow.
N O T E
Some filters, if not washed before use, release monomers that pollute the flow cell
windows and or cause air bubbles to stick in the flow cells.
3 Insert the probes into your dissolution vessels.
Attach the necessary fitting and tubing to the union, see
System Tubing"
page 49.
26
Probe for Dissolution Bath
on page 31, and
Agilent 8453 Dissolution Testing System Installing and Operating Manual
"Connecting Valve System Tubing"
"Connecting Multicell
on