6.3.2 Forced stop of auto calibration

This mode is used to cancel the auto calibration forcedly.
(1) During measurement, press the
ing the
key.
(2)
selection screen that appears, point the
pressing the
key.
ENT
(3)
A message appears, prompting you to
tinue auto calibration. To cancel the auto
calibration, press the
press the
ESC
stopped.
32
MODE
or the
key. Press the
or the
key. Press the
key. If you
ENT
key, auto calibration is not
key to
User Mode
-
ENT
Setting about Calibration
Alarm Setting
Setting of Auto Calibration
Changeover of Range
Parameter Setting
About Auto Cal.
Start Time
Cycle
Flow Time
ON/OFF
Stop Auto Calibrati
About Auto Cal.
-
Start Time
Cycle
Flow Time
ON/OFF
Stop Auto Calibrati
Select an item
with UP/DOWN and ENT
Select setting item
SUN 12 : 00
07
day
300 sec.
OFF
Time : MON 17 : 24
Stop auto calibration?
Stop with ENT
Not with ESC