Setting System Config - Chroma 19020 User Manual

Multi-channel sync hipot tester
Hide thumbs Also See for 19020:
Table of Contents

Advertisement

Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
User's Manual
The 19020 can set output channels CH1-CH10 while the 19020-4 can
only set channels CH1-CH4. The message line on the test screen will
show "Module Fail" if an incorrect channel is selected and the test will not
start.

4.4.3 Setting SYSTEM CONFIG

In the SYSTEM screen, move the highlighted cursor to SYSYEM CONFIG and press ENTER
to go to the SYSYEM CONFIG setting screen shown below:
SYSTEM CONFIG
01. CONTRAST
02. BEEPER
03. GPIB
04. RS232
05. LINK SETUP
06. LINK ADDRESS
07. ACA MEAS.
08. DC ARC RATE
1-16
In the SYSTEM CONFIG screen, press the ▲, ▼ keys to move the highlighted cursor to the
parameter to be set.
The following table lists the parameters in the SYSYTEM CONFIG menu and their
descriptions.
Parameter
Range
CONTRAST
1 - 16
BEEPER
HIGH / LOW / OFF
GPIB
UNINSTALLED /
ADDRESS = 0-30
RS232
9600 / 19200 / 38400
LINK SETUP
MASTER/SLAVE/SCA
NNER
LINK ADDRESS 1-9
ACA MEAS
RMS/GENERAL
DC ARC RATE 1.0-10.0
4-10
:
06
UP
:
LOW
: UNINSTALLED
DOWN
:
9600
:
MASTER
:
0
:
RMS
:
1.0
EXIT
REMOTE LOCK OFFSET ERROR
Default
Description
06
Adjusts the LCD brightness.
LOW
Adjusts the beeper volume.
03
Sets the GPIB interface address.
Shows UNINSTALLED if the GPIB
card is not installed.
9600
Sets the transmission baud rate of the
RS232 interface.
MASTER Sets the tester to MASTER or SLAVE
when linked for test.
Set to SCANNER when connecting to
A190201/A190202.
0
Sets the tester address when linked for
test.
Note:
The tester address is fixed at 0
when set to MASTER.
RMS
Sets the AC current measurement.
1.0
The ARC LIMIT setting will multiply this
rate for RAMP, DWELL time during DC
MODE test.

Advertisement