DCV 6kV FULL (4kV) Calibration Screen:
CALIBRATION
DCV 6kV FULL (4kV)
CHANNEL: 1
OUTPUT
4.000kV
STANDARD
3.991kV
DEF=4kV
1.
Press START to output voltage and read the data from the high voltage meter.
2.
Press the 0
. - 9 numeric keys to input the reading from the high voltage meter.
3.
Press ENTER to confirm the input.
4.
Press STOP to stop the high voltage output.
5.
Press ▼ to go to the OSC 100V OFFSET (50V) calibration.
7.4.3 OSCV Calibration
OSC 100V OFFSET(50V) Calibration Screen:
CALIBRATION
OSC 100V OFFSET (50V)
CHANNEL: 1
OUTPUT
50V
STANDARD
62V
DEF=50V
1.
Press START to output voltage and read the data from the high voltage meter.
2.
Press the 0
. - 9 numeric keys to input the reading from the high voltage meter.
3.
Press ENTER to confirm the input.
4.
Press STOP to stop the high voltage output.
5.
Press ▼ to go to the OSC 100V FULL (100V) calibration.
READING
3.982kV
REMOTE LOCK
READING
55V
REMOTE LOCK
Calibration Procedure
NEXT CH
PREV CH
CORR
ERROR
NEXT CH
PREV CH
CORR
ERROR