Setting Test Control - Chroma 19020 User Manual

Multi-channel sync hipot tester
Hide thumbs Also See for 19020:
Table of Contents

Advertisement

Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
User's Manual

4.4.2 Setting TEST CONTROL

In the SYSTEM screen, move the highlighted cursor to TEST CONTROL and press ENTER
to go to the TEST CONTROL setting screen shown below:
TEST CONTROL
01. PASS HOLD
02. ACV FREQUENCY
03. SOFTWARE AGC
04. MIN. VOLTAGE
05. WV AUTO RANGE
06. CH AFTER FAIL
07. RAMP JUDGMENT
08. DEF. CHANNELS
09. SCREEN
10. EOT
11. DISCH. Vmin
12. AC OFFSET
0.2-99.9S
In the TEST CONTROL screen, press the
parameter to be set.
The following table lists the parameters in the TEST CONTROL menu and their descriptions.
No.
Parameter
01 PASS HOLD
02 AC
FREQUENCY
08 SOFTWARE
AGC
04 MIN. VOLTAGE
05 WV AUTO
RANGE
06 CH AFTER FAIL
RAMP
07
JUDGMENT
DEF.
08
CHANNELS
09 SCREEN
10 EOT
4-8
:
:
:
:
:
:
:
:
:
:
:
:
REMOTE LOCK OFFSET ERROR
,
Range
Default
0.2-99.9sec
0.2sec
50, 60Hz
60
ON/OFF
ON
80%
OFF,50-95%
ON/OFF
OFF
STOP/
STOP
CONTINUE
ON / OFF
ON
Output
channels of
-------------
Frame 0 are
ON
ON/OFF
ON
TIMER /
TEST
0.2 sec
60 Hz
ON
80%
OFF
STOP
ON
STEUP
ON
TEST
OFF
EXIT
0.01
keys to move the highlighted cursor to the
Description
Sets the length of time the beeper
sounds for a PASS judgement.
Sets the frequency of the output voltage
when doing the AC withstand voltage
test.
Sets the software AGC function to ON or
OFF.
Sets the percentage for the output
voltage to reach the set voltage.
Sets the auto range function for the
withstand voltage test to ON or OFF.
Sets if the channel stops testing when a
FAIL occurs during a test.
When set to ON, determines if the current
readings exceed the high limit during
ramp time execution.
When set to OFF, does not determine if
the current readings exceed the high limit
during ramp time execution.
Sets the default channels to be ON. The
setting here will become the default
PROGRAM channel. See Setting DEF.
CHANNELS for details.
Turns the LCD screen ON or OFF during
a test.
1. When EOT is set to TIMER, nEOT