7.5.2 DCA Current Calibration

Connect a 10M
0.5 Watt or high power simulated load resistance between the
high voltage output terminal (HV OUTPUT) of the channel to be calibrated on this
tester and the high potential terminal of the AC meter. Connect the low potential
terminal (RTN/LOW) of the channel to be calibrated on this tester to the low
potential terminal of the AC meter.
DCA 3mA OFFSET(0.12mA) Calibration Screen:
CALIBRATION
DCA 3mA OFFSET(0.12mA)
CHANNEL: 1
OUTPUT
1.200kV
STANDARD
0.120mA
LOAD=10M
1.
Press START to output voltage and read the data from the DC meter.
2.
Press the 0
. - 9 numeric keys to input the reading from the DC meter.
3.
Press ENTER to confirm the input.
4.
Press STOP to stop the high voltage output.
5.
Press ▼ to go to the DCA 3mA FULL (2.4mA) calibration.
Change the simulated load resistance to 500k
DCA 3mA FULL(2.4mA) Calibration Screen:
CALIBRATION
DCA 3mA FULL(2.4mA)
CHANNEL: 1
OUTPUT
1.200kV
STANDARD
2.401mA
LOAD=500k
1.
Press START to output voltage and read the data from the DC meter.
2.
Press the 0
. - 9 numeric keys to input the reading from the DC meter.
READING
0.118mA
REMOTE LOCK
10watt or higher power.
READING
2.378mA
REMOTE LOCK
Calibration Procedure
NEXT CH
PREV CH
CORR
ERROR
NEXT CH
PREV CH
CORR
ERROR