Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
User's Manual

4.7.2 Using GET OFFSET

1.
Press the [OFFSET] Function Key in the TEST screen to go to the GET OFFSET screen
shown below:
GET OFFSET
1.PLEASE OPEN THE OUTPUT
TERMINAL.
PRESS <START> TO START
STANDBY
2.
GET OFFSET offsets the leakage current of AC / DC / IR MODE test leads and fixtures
as well as the stray capacitance of the OSC MODE.
3.
Remove the UUT from the fixture and press START to offset the leakage current or
leakage capacitance.
4.
The message box [CORR] will be highlighted when the offset has been completed.
5.
Pressing the [OFF] Function Key will cancel OFFSET.

4.7.3 Using GET Cs

1.
Press the [GET Cs] Function Key in the TEST screen to go to the GET Cs screen shown
below:
GET Cs
1.PLEASE CONNECTED THE
STANDARD UUT.
2.GET OFFSET BEFORE THAT.
3.ONLY FOR OSC MODE.
4.Cs VALUE WILL BE REPLEASED.
PRESS <START> TO START
STANDBY
2.
Use the standard sample under the capacitance test as the UUT and connect it to the 1
channel opened by OSC (this CHANNEL has to be the Master CHANNEL). Press
to get the standard capacitance (GET Cs).
START
4-24
OFF
EXIT
REMOTE LOCK
CORR
EXIT
REMOTE LOCK
CORR
ERROR