Arc Calibration; Ac Arc Calibration - Chroma 19020 User Manual

Multi-channel sync hipot tester
Hide thumbs Also See for 19020:
Table of Contents

Advertisement

Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
User's Manual
1.
Press START to output voltage.
2.
Press the 0
. - 9 numeric keys to input the actual resistance.
3.
Press ENTER to confirm the input.
4.
Press STOP to stop the high voltage output.
5.
Press ▼ to go to the AC ARC 15mA (7mA) calibration.

7.7 ARC Calibration

1.
WARNING
2.

7.7.1 AC ARC Calibration

Connect one end of a 250k
the high voltage output terminal (HV OUTPUT) of the channel to be calibrated on
the tester, and move the other end close to the low potential terminal (RTN/LOW) of
the channel to be calibrated without making any physical connection in order to
create sparks.
AC ARC 15mA (5mA) Calibration Screen:
CALIBRATION
AC ARC 15mA(5mA)
CHANNEL: 1
OUTPUT
1.250kV
STANDARD
5mA
LOAD=500k
1.
Press START to output voltage.
2.
When the status indicator shows PASS the device has returned the correct ARC value
and turned off the high voltage output.
3.
Press ▼ to go to the DC ARC 5mA (3mA) calibration.
7-22
ARC calibration is a dangerous test as the high voltage is exposed
outside the terminal. Use extreme caution when performing the test.
For more detailed information, contact Chroma or its local
distributors.
10Watt or higher power simulated load resistance to
READING
4.1mA
REMOTE LOCK
NEXT CH
PREV CH
CORR
ERROR