7.7.2 DC ARC Calibration
DC ARC 5mA (3mA) Calibration Screen:
CALIBRATION
DC ARC 5mA(3mA)
CHANNEL: 1
OUTPUT
1.500kV
STANDARD
3.0mA
LOAD=500k
1.
Press START to output voltage.
2.
When the status indicator shows PASS the device has returned the correct ARC value
and turned off the high voltage output.
7.8 When Calibration is Done
1.
Press SYSTEM to exit the calibration mode or press ▼ to return to the A/D
REFERENCE calibration screen.
2.
Press the [NEXT CH],[PREV CH] Function Keys to select another channel for
calibration.
READING
1.9mA
REMOTE LOCK
Calibration Procedure
NEXT CH
PREV CH
CORR
ERROR
7-23