4.4.3 Setting SYSTEM CONFIG

In SYSTEM screen, move the highlight to SYSYEM CONFIG and ENTER to go to SYSYEM
CONFIG setting screen as shown below:
SYSTEM CONFIG
01. CONTRAST
02. BEEPER
03. GPIB
04. RS232
05. LINK SETUP
06. LINK ADDRESS
07. ACA MEAS.
08. DC ARC RATE
1-16
When in SYSTEM CONFIG screen, press ▲, ▼ to move the highlight to the item desired for
setting the related function.
The following table lists the setting items of SYSYTEM CONFIG and their descriptions.
Setting Items
Range
CONTRAST
1 - 16
BEEPER
HIGH / LOW / OFF
GPIB
UNINSTALLED /
ADDRESS = 0~30
RS232
9600 / 19200 / 38400
LINK SETUP
MASTER/SLAVE/SCA
NNER
LINK ADDRESS 1~9
ACA MEAS
RMS/GENERAL
DC ARC RATE 1.0~10.0
1.
2.
4-10
:
:
: UNINSTALLED
:
:
:
:
:
REMOTE LOCK OFFSET ERROR
Default
06
LOW
03
9600
MASTER It sets the tester to be MASTER or
0
RMS
1.0
When LINK SETUP is set to SLAVE, besides the items "02.
SYSTEM CONFIG", "05. CHANGE PASSWORD", "06.
CALIBRATION", "07. ERROR LOG" and "08. ABOUT" in the
SYSTEM screen, the rest items are all invalid.
Only the 19020/19020-04 LINK SETUP settings has SCANNER
selection.
06
UP
LOW
DOWN
9600
MASTER
0
RMS
1.0
EXIT
Description
It adjusts the LCD brightness.
It adjusts the beeper volume.
It sets the GPIB interface address. It
shows UNINSTALLED if GPIB card is
not installed.
It sets the transmission baud rate of
RS232 interface.
SLAVE when linked for test.
Be sure to set it to SCANNER when
connecting to A190201/A190202.
It sets the tester address when linked
for test.
Note:
The tester address is fixed to 0
when set to MASTER.
It sets the AC current measurement.
The ARC LIMIT setting will multiply this
rate for RAMP, DWELL time during DC
MODE test.