8.1 Changing rechargeable batteries /
batteries
Switch off the analyser before opening it!
Connect the analyser to the mains unit (accessories),
otherwise the analyser settings (e.g. date/time ...) and the
readings retained in the analyser are lost.
Saved Readings (
1 Place analyser on its front and open and remove battery
compartment cover at the back of the analyser.
2 Remove spent rechargeable batteries/batteries from the
battery compartment.
3 Place new rechargeable batteries/batteries in the battery
compartment. Insert correctly (+/-).
-
The analyser switches off once the rechargeable batteries /
batteries are inserted!
4 Close cover on battery compartment.
Changing the rechargeable battery block
The rechargeable battery block (
changed in the same way as the rechargeable
batteries/batteries, however it must be connected to the white
socket in the battery compartment via the plug-in connection
and held in position in the battery compartment via the
available adapter

8.2 Emptying the condensate trap

Switch off analyser before emptying the condensate trap!
1 Hold analyser with the condensate trap outlet in an upward
position and remove the plugs
opening.
The condensate is made up a weak acidic mixture. Avoid contact
with skin.
2 Empty condensate into container.
3 Close off outlet using plugs.