Pannoramic MIDI II 2.0.5 User's Guide
Click the
button to display barcode label area. To rotate the label area by 180°, click
– Undo / Redo modification made during specifying the Scanmap. These buttons are inactive
when Use threshold function is applied.
5.1.2.2 Focus Indicator
Click
to activate the Focus indicator function. Search for the sharpest focus on the current field
of view automatically by clicking
To define focus limits, select Limit the range. Set the upper and lower limits by selecting the desired
value in the selection boxes. By using Set +-Range, the Lower limit value is set to -32 of the current
focus value and the Upper limit value is set to +32 of the current focus value. This is useful if Auto
Focus does not find the focus on the specimen, but on the coverslip. Also, this makes the focusing
sequence faster.
For manual focusing define the manual focus step size by adjusting the Step size slider and use the
two arrow buttons to focus manually. This is useful when you are interested in an area that remained
blurry with Auto Focus (for example, the specimen is thicker in that area). This does not influence the
way slides are digitized in Auto focus mode.
December 7, 2018 – Rev. 1
Figure 33 – Barcode label area window
- Auto focus button.
Figure 34 – Focus indicator panel
3DHISTECH Ltd.
5.1.2 Microscope Control Panel
.
60(144)