Pannoramic MIDI II 2.0.5 User's Guide
7. Switch on the 24 V power supply.
8. Switch on the control computer.
9. Launch control software.
Virtual slide is improperly focused almost across its entire surface
Focus range enabled
•
Disable Limiting focus range function
◦
Poor image quality
Imaging error (for example, chromatic aberrations)
•
Contact 3DHISTECH Service for support.
◦
Do not modify the camera driver settings, as the default factory
settings give the most accurate images.
Virtual slides show poor stitching in Pannoramic Viewer
Scan camera is not properly adjusted (for example, due to loose fixture)
•
Contact 3DHISTECH Service for support.
◦
Slide is often positioned incorrectly
Guiding fault in transport system
•
Contact 3DHISTECH Service for support.
◦
Digitization process is extremely slow
Extended focusing option is activated
•
Deactivate the Extended focusing option if not required.
◦
Z-stack option is activated
•
Deactivate the Z-stack option if not required.
◦
December 7, 2018 - Rev. 1
Warning!
3DHISTECH Ltd.
7 Troubleshooting and Maintenance
126(144)