Pannoramic MIDI II 2.0.5 User's Guide
Do NOT use Remove coverslip function on TMA slides! The layout of
TMA slides may mislead the detecting algorithm.
Fill holes inside scanmap
Often these holes contain small sample objects that otherwise would not be scanned.
Remove specks
predetermined size. This helps filtering specks and other uninteresting objects out.
Set the maximum volume of specks manually in the setting window
Threshold settings have two different statuses:
Auto threshold
Make sure that the detection result is acceptable for you. Otherwise
set the threshold value manually.
Make sure that the detection result is acceptable for you. Otherwise set the threshold value
manually.
Manual threshold
preview with Calculate Threshold
manually. This sets the sensitivity of the specimen detection. To set the desired threshold
value manually, change the setting in the Specimen threshold selection box in the range of 0-
255. The default value is 212. Smaller values result in a more sensitive scanning, meaning
that more areas will be digitized, however, this may also include specks or other
imperfections of the slide. Greater values result in a less sensitive scanning, but it is possible
that some of the smaller specimen areas will not be digitized.
December 7, 2018 – Rev. 1
Warning!
: If selected, the holes inside the specimen will also be scanned.
: If selected, the scanner does not digitize objects smaller than the
: It automatically detects the specimen in the preview image.
Caution!
: Set threshold value manually, or calculate value on the actual
3DHISTECH Ltd.
function which is active only if setting the value