Pannoramic MIDI II 2.0.5 User's Guide
Lumencor SOLA SM II illumination unit
Keep inflammable and combustible materials outside the range of SOLA SM II illumination
unit.
Do not expose yourself to the UV radiation of the SOLA SM II illumination unit.
When the SOLA SM II illumination unit is switched on:
NEVER look directly into the fiber-optic cable.
NEVER remove the fiber-optic cable from the device.
NEVER connect the fiber-optic cable to the device.
NEVER remove the housing from the device.

7.1.1 Cleaning the device

Before cleaning, switch off the device and pull the plug out.
Do not let liquid penetrate into the system.
To clean liquid from the device, use a lint-free cotton cloth. You can moisten the cloth with
water.
To clean greasy surfaces, it is recommended to apply soapy water on a piece of lint-free
cotton cloth and gently rub the surface.

7.1.2 Cleaning the trays

NEVER clean the trays with Xylene, because it might harm the plastic. We recommend to apply
Isopropanol on a piece of lint-free cotton cloth and gently rub the surface instead. In case the tray is
damaged, use either order a new one, as inserting a deteriorated tray could lead into unsafe loading
of slides.

7.1.3 Preventive maintenance

To prevent fungus growth
Do not leave the equipment in an unconditioned environment with a storage temperature
above 60°C (140°F) or below 0°C (32°F), which may damage the equipment.
Maximum operating temperature is 35°C (95°F) with a relative humidity of less than 75%.
Install it in a room with proper air ventilation.
Do not let dust or fingerprints remain on optical surfaces.
December 7, 2018 - Rev. 1
3DHISTECH Ltd.
7 Troubleshooting and Maintenance
124(144)