Image is out of focus (in Manual mode)
Focus range limit is activated
•
Deactivate the Limiting focus range option.
◦
Focus range setting is incorrect
•
Change focus range setting.
◦
All individual images show a shading effect
Algorithm failed to find empty image fields
•
Capture a reference image with a blank slide.
◦
7.2.1 Fluorescent Applications (optional)
The switching of the Lumencor SPECTRA 6 cannot be triggered, in spite of trigger connection
through the camera.
Contact 3DHISTECH Service for support.
•
No suitable exposure time can be found. Signal is too weak.
Lumencor SOLA SM II is switched off
•
•
December 7, 2018 - Rev. 1
Switch light source on.