Change Of Set Flow And Hysteresis [Simple Setting Mode]; Simple Setting Mode - SMC Networks PF3A801H-L Operation Manual

Digital flow switch modular type
Hide thumbs Also See for PF3A801H-L:
Table of Contents

Advertisement

Change of Set Flow and Hysteresis [Simple setting mode]

■Simple setting mode

In the simple setting mode, the set value, hysteresis, and delay time can be changed while checking the
current measured value (main display).
<Operation>
[Hysteresis mode]
(1) Press the SET button for 1 second or longer (but less than 3 seconds) in measurement mode. [SEt] is
displayed on the main display.
When the button is released while in the [SEt] display, the current measured value is displayed on the
main display, [P_1]/[P_2] or [n_1]/[n_2] is displayed on the sub display (left) and the set value is
displayed on the sub display (right).
(2) Change the set value using the UP or DOWN button, and press the SET button to set the value. Then,
the setting moves to hysteresis setting.
(The snap shot function can be used. (Refer to page 69))
(3) Change the set value using the UP or DOWN button, and press the SET button to set the value. Then,
the setting moves to the setting of OUT2.
(The snap shot function can be used. (Refer to page 69))
(4) Like the setting of OUT1, the setting returns to the setting of OUT2 by pressing the SET button after
setting the set value and hysteresis.
(5) Press and hold the SET button for 2 seconds or longer to complete the simple setting.
(If the button is pressed for less than 2 seconds, the setting will be returned to [P_1].)
1: Selected items of (1) to (4) become valid after pressing the SET button.
2: After enabling the setting by pressing the SET button, it is possible to return to measurement mode by pressing the SET
button for 2 seconds or longer.
3: When the setting target is set to accumulated pulse, error output or output OFF (refer to page 38), the simple setting mode
cannot be used. (the setting returns to measurement mode by releasing the button when [SEt] is displayed.)
: When OUT1 and OUT2 are set to accumulated pulse
output, error output or output OFF [---] will be
displayed in the sub screen when [SEt] is displayed.
It is not possible to move to the Simple setting mode.
-29-