F9, F9 Express Fetal & Maternal Monitor User Manual
A3.9 NIBP and SpO2 No Results
Phenomenon
NIBP and SpO
have no
2
results
A3.10 Blown Fuses
Switch off the monitor and unplug it before changing the fuse.
Replace the fuse when it is blown.
The two fuses of the monitor are located on the bottom panel, their specifications are:
Size: Ф5mm*20mm; Model: T2AH250V.
To replace a fuse:
Fold the LCD display completely flat.
1)
Carefully place the monitor upside down on a flat surface covered with cloth or other
2)
protecting pad.
With a flat-head screw driver, push the fuse in for about 1 mm and then unscrew it
3)
anticlockwise.
Remove the old fuse and replace it with a new fuse that is supplied by the manufacturer or of
4)
the same specifications.
Push the new fuse into the socket for about 1 mm and then screw it clockwise back in
5)
position.
The
special
ground
connecting with monitor is not
properly earthed.
Possible Cause
The NIBP cuff is not properly
wrapped
to
the
position
patient's arm.
The NIBP can not be inflated.
Hose connector plug is not
connected well with the NIBP
socket.
SpO
transducer
2
connected well with the SpO
socket.
Abnormal working condition.
WARNING
- 175 -
wire
Check if the special ground wire
connecting
earthed.
Check if the NIBP cuff is
of
properly wrapped to the position
of patient's arm.
Extend catheter, and check the
connection.
Check if the hose connector
plug is connected well with the
NIBP socket.
is
not
Check if the SpO
connected well with the SpO
2
socket.
Shut off the power, then switch it
on again.
Troubleshooting
with
monitor
is
Solution
transducer is
2
2