EDAN F9 User Manual page 207

Fetal & maternal monitor
Hide thumbs Also See for F9:
Table of Contents

Advertisement

F9, F9 Express Fetal & Maternal Monitor User Manual
Voltage dips, short
interruptions
and
voltage
variations
on
power
supply
input lines
IEC 61000-4-11
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
0 % U
0,5 cycle
T;
At 0° , 45° , 90° , 135° ,
180° , 225° , 270° and
315°
0 % U
; 1 cycle
T
and
70
%
U
;
25/30
T
cycles )
Single phase: at 0°
0 % U
; 250/300
T
cycle
0 % U
0,5 cycle
T;
At 0° , 45° , 90° , 135° ,
180° , 225° , 270° and
315°
0 % U
; 1 cycle
T
and
70
%
U
;
25/30
T
cycles )
Single phase: at 0°
0 % U
; 250/300
T
cycle
- 196 -
EMC Information
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the F9 and F9 Express Fetal
& Maternal Monitors requires
continued
operation
power mains interruptions, it is
recommended that the F9 and
F9 Express Fetal & Maternal
Monitors be powered from an
uninterruptible power supply or
a battery.
during