F9, F9 Express Fetal & Maternal Monitor User Manual
For network safety:
63 Given the vulnerabilities of WEP, it is advised to use WAP and WAP2. WEP should
not be used as it is not secure.
64 Data in transit is no encrypted. Local Area Network isolation or other protections are
suggested to prevent unintended exposure of PHI.
In addition, when you use the FTS-3 fetal telemetry system, please pay attention to the
warnings as follows:
65 The system should be operated by the doctor or under the doctor's instructions.
66 SHOCK HAZARD – The base station and transducers for one patient must be
supplied by the same power and do not change the power supply.
67 Please arrange a function test periodically for the system.
68 Do not move the system when it is powered on and do not soak it in any liquid.
69 Please check the transducer, cable and base station periodically. If the transducers
are damaged, do not use them in water and repair them in time.
70 The battery in the wireless transducer should be replaced by the serviceman
authorized by EDAN.
71 If the transducer has been beaten or knocked, please check whether the cover is
airproof or damaged. If you have any doubt, please contact the manufacturer or local
agent.
72 If the battery in the base station is stored alone and not used for a long time, we
recommend that the battery should be charged at least once every 6 months to
prevent over discharge.
73 The wireless transducer has priority over the wired transducer. When the wireless
transducer is working, the wired transducer will be turned off automatically. Do not
use the wireless transducer and the wired transducer at the same time.
1
The device is designed for continuous operation. Avoid liquid splashing on the
device.
2
Refer servicing to qualified personnel.
3
Keep the environment clean. Avoid vibration. Keep it far from corrosive medicine,
dust area, high-temperature and humid environment.
4
When installing the unit into a cabinet, allow for adequate ventilation, accessibility for
servicing, and room for adequate visualization and operation.
5
Do not operate the unit if it is damp or wet because of condensation or spills. Avoid
using the equipment immediately after moving it from a cold environment to a warm,
humid location.
6
The appliance coupler or mains plug is used as isolation means from supply mains.
Position the monitor in a location where the operator can easily access the
disconnection device.
WARNING
CAUTION
- 8 -
Safety Guidance