EDAN F15 Series Service Manual
Hide thumbs Also See for F15 Series:
Table of Contents

Advertisement

Advertisement

Table of Contents
loading

Summary of Contents for EDAN F15 Series

  • Page 2 This manual will help you understand the operation and maintenance of the product better. It is reminded that the product shall be used strictly complying with this manual. User’s operation failing to comply with this manual may result in malfunction or accident for which Edan Instruments, Inc. (hereinafter called EDAN) can not be held liable.
  • Page 3 injury or death. CAUTION A CAUTION label advises against actions or situations that could damage equipment, produce inaccurate data, or invalidate a procedure. NOTE A NOTE provides useful information regarding a function or a procedure.
  • Page 4: Table Of Contents

    Content Chapter 1 Safety Guidance ..........................7 1.1 Introduction ..............................7 1.2 General Information ............................ 7 1.3 Safety Precautions ............................8 1.4 Definitions and Symbols ..........................1 Chapter 2 Installation ............................5 2.1 Environmental Requirements ........................5 2.2 Electrical Requirements ..........................6 2.3 Safety Requirements ............................
  • Page 5 8.3 Verifying Malfunction of Lower Case Interface Board ................43 8.4 Verifying Malfunction of AC-DC Power Board ..................44 8.5 Verifying Malfunction of EDAN Transducer Socket Board (wired model only) ........45 8.6 Verifying Malfunction of NELL Transducer Socket Board (wired model only) ........46 8.7 Verifying Malfunction of Wireless Charging Board ..................
  • Page 6 10.2.4 SpO2 Performance Test ........................62 10.2.5 NIBP Performance Test ........................62 10.2.6 TEMP Performance Test ........................62 10.2.7 FECG Performance Test(Wireless Model Only) ................. 63 Appendix 1 Renewal Parts ..........................65 Appendix 2 Connecting to OB TraceVue System .................... 66...
  • Page 7: Chapter 1 Safety Guidance

    F15 Series Fetal & Maternal Monitor Service Manual Safety Guidance Chapter 1 Safety Guidance NOTE: This service manual is written to cover the maximum configuration. Therefore, your model may not have some of the parameters/modules described, depending on what you have ordered.
  • Page 8: Safety Precautions

    F15 Series Fetal & Maternal Monitor Service Manual Safety Guidance does not accept responsibility for safety compliance, reliability and performance if modifications or repairs are carried out by unauthorized personnel. Identical replacement parts must be used.  The typical operator's position is in front of the monitor. Please position the device in a location where the operator can easily see the screen and access the operating controls.
  • Page 9 F15 Series Fetal & Maternal Monitor Service Manual Safety Guidance WARNING The monitor must be serviced by authorized and qualified personnel only. EDAN do not accept responsibility for safety compliance, reliability and performance if modifications or repairs are carried out by unauthorized personnel. Identical replacement parts must be used.
  • Page 10 F15 Series Fetal & Maternal Monitor Service Manual Safety Guidance WARNING 12 SHOCK HAZARD - Don’t connect electrical equipment, which has not been supplied as a part of the system, to the multiple portable socket-outlets supplying the system. 13 Do not use the additional multiple portable socket-outlet or extension cord in the medical electrical system, unless it’s specified as part of the system by manufacturer.
  • Page 11 F15 Series Fetal & Maternal Monitor Service Manual Safety Guidance WARNING 24 Parts and accessories used must meet the requirements of the applicable IEC 601 series safety standards, and/or the system configuration must meet the requirements of the IEC 60601-1 medical electrical systems standard.
  • Page 12: Definitions And Symbols

    F15 Series Fetal & Maternal Monitor Service Manual Safety Guidance 1.4 Definitions and Symbols F15 Series Fetal & Maternal Monitor Fetal monitoring socket, ultrasound transducers, TOCO transducer and Fetal&Maternal Module can be connected to it( Type CF applied part) Socket for Remote Event Marker ( Type CF applied part)
  • Page 13 F15 Series Fetal & Maternal Monitor Service Manual Safety Guidance Alternating Current Direct Current Stand-by Power Supply switch Warning (Background: Yellow; Symbol & outline: Black) Caution Operating instructions Refer to User Manual (Background: Blue; Symbol: White) TYPE CF APPLIED PART...
  • Page 14 F15 Series Fetal & Maternal Monitor Service Manual Safety Guidance Manufacturer AUTHORISED REPRESENTATIVE IN THE EUROPEAN COMMUNITY Disposal method Non-ionizing electromagnetic radiation USB Port Ethernet Port Serial Number Catalogue number Federal (U.S.) Law restricts this device to sale by or on the order...
  • Page 15 F15 Series Fetal & Maternal Monitor Service Manual Safety Guidance DO NOT STEP ON - 4 -...
  • Page 16: Chapter 2 Installation

    F15 Series Fetal & Maternal Monitor Service Manual Installation Chapter 2 Installation WARNING The system installation should be operated by serviceman authorized by the manufacturer. 2.1 Environmental Requirements Working Environment 0 º C ~ + 40 º C ( +32 º F ~ +104 º F)
  • Page 17: Electrical Requirements

    F15 Series Fetal & Maternal Monitor Service Manual Installation EMC: Group I Class A Working System Continuous running equipment  If the monitor is installed in a cabinet, allow at least 2 inches (5 cm) clearance around the monitor for proper air circulation; allow adequate accessibility for servicing, and adequate room for visualization and operation.
  • Page 18: Mounting The Monitor On A Wall

    F15 Series Fetal & Maternal Monitor Service Manual Installation 2.4.2 Mounting the Monitor on a Wall CAUTION 1 The monitor should only be mounted on a solid concrete or brick wall. 2 Shut the display completely flat before mounting the monitor to the wall.
  • Page 19: Installing The Monitor On Trolley

    F15 Series Fetal & Maternal Monitor Service Manual Installation Wall Wall Mounting Board Figure 2-2 Fixing Wall Mounting Board Lift the monitor with the bottom panel facing the wall. Let the four posts on the board stretch into wall-mounting holes and then release your hands from the monitor slowly.
  • Page 20: Connecting To Ac Power

    F15 Series Fetal & Maternal Monitor Service Manual Installation Figure 2-3 the Monitor on a Trolley (Sketch Map) 2.5 Connecting to AC Power Apply the power cable provided with the monitor. Plug one end of the power cable to the power socket of the monitor.
  • Page 21: Chapter 3 Functional Checks

    F15 Series Fetal & Maternal Monitor Service Manual Functional Checks Chapter 3 Functional Checks This section describes the procedure of a complete functional test to support recommended preventive-maintenance schedules. You are not required to open the device case for functional checks.
  • Page 22: Printing Check

    F15 Series Fetal & Maternal Monitor Service Manual Functional Checks 3.4 Printing Check Press the PRINT key to start printing. Check if the recorder starts printing. Enable the Print Self-Test function. Restart the monitor and verify if the recorder prints a baseline.
  • Page 23: Chapter 4 System Configuration

    F15 Series Fetal & Maternal Monitor Service Manual System Configuration Chapter 4 System Configuration The end users can not change the system configurations of the monitor. As a service engineer, you need to change these configurations for them after the monitor is installed and checked properly.
  • Page 24: Clearing Data

    F15 Series Fetal & Maternal Monitor Service Manual System Configuration 4.4 Clearing Data The auto-saved data is saved in the monitor memory. In order to delete this data, 1 Open the System Setup menu. 2 Select Clear Data. 3 Select OK.
  • Page 25: Changing The Device No

    F15 Series Fetal & Maternal Monitor Service Manual System Configuration up/left/right key and ENTER key to enter the calibrating interface. If the touching function still fails or calibration cannot be performed, please refer to section 7.3 for service. 4.7 Changing the Device No.
  • Page 26: Configuring For Ethernet Connecting

    F15 Series Fetal & Maternal Monitor Service Manual System Configuration channel number and wireless module number: 1 Open the System Setup menu. 2 Select Network Setup. 3 Select Wireless CH. 4 Select a number (6 ~ 31). 5 Select Wireless MOD.
  • Page 27: Changing Trace Width

    F15 Series Fetal & Maternal Monitor Service Manual System Configuration 192.168.1.X. 14 Select OK in the Network Setup menu. 15 Select OK in the System Setup menu and then restart the monitor. 4.11 Changing Trace Width In order to differentiate the traces clearly, you may assign different widths for different traces.
  • Page 28: Changing Fhr Algorithm

    F15 Series Fetal & Maternal Monitor Service Manual System Configuration 4.14 Changing FHR Algorithm The Beat-to-beat FHR algorithm provides vivid short-term variation information of the fetal heart rate. The Average FHR algorithm provides a smoothened FHR trace. 1 Open the System Setup menu.
  • Page 29 F15 Series Fetal & Maternal Monitor Service Manual System Configuration • T-fitting • NIBP extension tubes • Cylinder (200 ml) • Manometer (Its measurement range should be within the range of 0 mmHg to 300 mmHg; its accuracy should be more precise than the accuracy of ± 0.3 mmHg.)
  • Page 30: Chapter 5 Maintenance

     Check all the outer cables, power socket and power cables.  Check if the monitor functions properly. If any damage is detected, replace the damage part(s) or contact EDAN immediately. (2) Routine Inspection The overall check of the monitor, including the safety check and functional check, should be performed by qualified personnel every 6 to 12 months, and each time after service.
  • Page 31: Maintenance Of The Transducers

    3) Remove the dust from the paper sensing mechanism and paper sensor with a dry clean cloth. CAUTION Only use the recorder paper provided by EDAN, or it may damage the recorder. This kind of damage is not covered by warranty.
  • Page 32: Chapter 6 Principle Introduction

    Chapter 6 Principle Introduction 6.1 System Principle Block Diagram F15 series fetal&maternal monitor consists of the AC-DC power board, DC-DC transfer board, battery, fan, main control board, thermal printer, display/touching module, bottom interface board, loudspeaker, pump&valve assembly, iM X12 parameter board, transducer socket board and wired transducer/all parameter accessories.
  • Page 33: Main Control Board

    F15 Series Fetal & Maternal Monitor Service Manual Principle Introduction Bottom Interface Board AC-DC Power Board eles dspe aker rgin Main Control Board DC-DC Transfer ssio Board trol Wireless Batt transducer/ Thermal Printer NIBP&Spo 2 module Display/touching Module Figure 6-1. F15 System principle block diagram 6.1.1 Main Control Board...
  • Page 34: Bottom Interface Board

    F15 Series Fetal & Maternal Monitor Service Manual Principle Introduction Serial Port 8 serial port ways Touch Screen Interface 1 interface. 1 standard USB HOST interface, 1 standard DEVICE interface (1.1 specification) HDMI 1 HDMI output port 1920*1080@60Hz interface, (highest resolution) Loudspeaker Interface 1 interface.
  • Page 35: Ac-Dc Power Board

    F15 Series Fetal & Maternal Monitor Service Manual Principle Introduction RJ45 Isolation RJ45 Transformer Interface Circuit HDMI Interface HDMI Protective Interface Circuit Socket Communication DB15 Interface Interface Circuit connect with the main USB HUB USB Interface control board RF Module...
  • Page 36: Dc-Dc Transfer Board

    F15 Series Fetal & Maternal Monitor Service Manual Principle Introduction 100V-240VAC Secondar Direct Input y Rectifier Correcti Changin Current Rectifier Filter 19.3V Filter Controll Controll Optoc oupler L6563S L6599A Figure 6-3 AC-DC Power Board Principle Block Diagram The parameters of this power board are listed below:...
  • Page 37: Im X12 Parameter Board(Wired Model Only

    X12 parameter board is located on the left of the main control board of F15 wired model, providing parameters like NIBP, SpO2(EDAN or Nellcor), TEMP. The ECG function of the board is not used. Its main specifications are as below.
  • Page 38 F15 Series Fetal & Maternal Monitor Service Manual Principle Introduction Measurement Time Span 120 Seconds (Maximum) Software Overpressure Protection (297 ± 3) mmHg Hardware Overpressure Protection (320 ± 10)mmHg [(41.3± 1.3) kPa] (2)SpO2(EDAN) Measurement Range 50% ~ 100% Resolution 70% ~ 100%: ± 2% Accuracy 50% ~ 69%:...
  • Page 39: Edan Transducer Socket Board (Wired Model Only)

    Every 1 ~ 2 second(s) 6.1.6 Edan Transducer Socket Board (wired model only) The EDAN transducer socket board locates on the left of the lower case, i.e. the left of the monitor. It communicates between the main control board and external transducer accessories. Its function is to isolate the signal of accessories, power supply and ground, and transfers collected SpO2 and TEMP signals to iM X12 parameter board.
  • Page 40: Wireless Charging Board (Wireless Model Only)

    F15 Series Fetal & Maternal Monitor Service Manual Principle Introduction Transducer Communicate interface*4(any Optocoupler signal fetal transducer at Socket to any socket connect with lower case interface board Optocoupler MARK_ISO、 MARK、EXT EXT_ISO +5V_I Supply power Isolation +5V5 for transducer Circuit...
  • Page 41: Interface

    F15 Series Fetal & Maternal Monitor Service Manual Principle Introduction Wireless Charging Controller Wireless Wireless Power Identificatio Charging Charging Analog Transmitter Supply n Controller Front End Coil NFC Controller Transmitter Coil Figure 6-7 Wireless Charging Board Principle Block Diagram 6.2 Interface On the rear panel of the monitor, there are two interfaces and an antenna: ...
  • Page 42: Hdmi Interface

    F15 Series Fetal & Maternal Monitor Service Manual Principle Introduction 6.2.4 HDMI Interface All the monitors are configured with a HDMI interface. It is used to project the display of the monitor to an external display, the output resolution is 1080P.
  • Page 43: Chapter 7 Troubleshooting

    It can also help you to have more exact descriptions of the fault symptoms when calling for service. EDAN supports replacement of PCBs and major subassemblies for this product. Verify the malfunction of the PCBAs/Modules prior to disassembling with the method described in chapter 8 Modules’...
  • Page 44: Touch Screen Failures

    F15 Series Fetal & Maternal Monitor Service Manual Troubleshooting displays improper LCD defective. Replace the LCD. characters. 7.3 Touch Screen Failures Phenomenon Possible Cause Solution Check that the touch screen related The touch screen related cables between the main control...
  • Page 45: Network Failures

    F15 Series Fetal & Maternal Monitor Service Manual Troubleshooting The paper drawer is not Push the drawer back in position. closed properly. Paper correctly Reload paper correctly. loaded. The printhead adjusting nuts Adjust the printhead adjusting nuts. Blank, blurry or incorrect are lopsided.
  • Page 46: Alarm Failures

    F15 Series Fetal & Maternal Monitor Service Manual Troubleshooting 7.8 Alarm Failures Phenomenon Possible Cause Solution FETUS 302 MODULE MALF Wired and wireless model is Re-set the model type. selected by mistake. Extend 302 MODULE MALF FETUS Main control board defective.
  • Page 47: Spo2 Monitoring Failures

    F15 Series Fetal & Maternal Monitor Service Manual Troubleshooting connection Check the connection. electrodes. TOCO&MECG Transducer Replace TOCO&MECG defective. Transducer. Main control board Replace the main control board. defective. 7.11 SpO2 Monitoring Failures Phenomenon Possible Cause Solution transducer defective. Replace the SpO transducer.
  • Page 48: Temp Monitoring Failures

    F15 Series Fetal & Maternal Monitor Service Manual Troubleshooting 7.13 TEMP Monitoring Failures Phenomenon Possible Cause Solution TEMP transducer defective. Replace the TEMP transducer. Bad connection of TEMP Check the connection. transducer. No TEMP numeric value. iM X12 parameter board Replace the iM X12 parameter board.
  • Page 49: Chapter 8 Modules' Malfunction Verification

    Modules’ Malfunction Verification F15 Series Fetal & Maternal Monitor Service Manual Chapter 8 Modules’ Malfunction Verification When a module is suspected of malfunction, verify it with the method described in this chapter. NOTE: For all the interfaces, the first pin is the one with a square solder pad.
  • Page 50 Modules’ Malfunction Verification F15 Series Fetal & Maternal Monitor Service Manual 1.425V@1GHz ) TP60 3.14~3.46V DC-DC power output on the main control board TP141 3.14~3.46V TP142 3.14~3.46V TP53 3.14~3.46V TP313 5.04~5.56V TP40 4.75~5.25V TP298 11.40~12.60V - 39 -...
  • Page 51 Modules’ Malfunction Verification F15 Series Fetal & Maternal Monitor Service Manual TP61 ⑧ Main control board - 40 -...
  • Page 52 Modules’ Malfunction Verification F15 Series Fetal & Maternal Monitor Service Manual TP130 TP142 TP53 TP128 TP129 TP40 TP298 TP60 TP131 TP198 TP25 TP26 TP19 TP20 TP17 TP18 TP16 TP313 TP22 TP141 Back of the Main control board 4) Compare the measurement results with the reference results in the list. If any one of the results exceeds the reference range, the main control board defection is confirmed.
  • Page 53: Verifying Malfunction Of The Im X12 Parameter Board

    Modules’ Malfunction Verification F15 Series Fetal & Maternal Monitor Service Manual 8.2 Verifying Malfunction of the iM X12 Parameter board To verify the iM X12 parameter board, Open the main unit using the procedures described in section 9.3. Switch on the monitor. (Powered by AC or battery.)
  • Page 54: Verifying Malfunction Of Lower Case Interface Board

    Modules’ Malfunction Verification F15 Series Fetal & Maternal Monitor Service Manual TP17 TP15 TP16 If any one of the results exceeds the reference range, the iM X12 parameter board defection is confirmed. Replacement of the iM X12 parameter board is recommended.
  • Page 55: Verifying Malfunction Of Ac-Dc Power Board

    Modules’ Malfunction Verification F15 Series Fetal & Maternal Monitor Service Manual ○ TP16 ○ TP52 ○ TP32 ○ TP31 ○ ○ TP18 ○ TP16 ○ TP18 Lower case interface board If any one of the results exceeds the reference range, the lower case interface board defection is confirmed.
  • Page 56: Verifying Malfunction Of Edan Transducer Socket Board (Wired Model Only)

    Symbol ○ TP23 ○ TP28 ○ TP27 ○ TP26 EDAN Transducer Socket Board If any one of the results exceeds the reference range, the EDAN transducer socket board defection is confirmed. Replacement of the board is recommended. - 45 -...
  • Page 57: Verifying Malfunction Of Nell Transducer Socket Board (Wired Model Only)

    Modules’ Malfunction Verification F15 Series Fetal & Maternal Monitor Service Manual 8.6 Verifying Malfunction of NELL Transducer Socket Board (wired model only) To verify the NELL transducer socket board, Open the main unit using the procedures described in section 9.3.
  • Page 58 Modules’ Malfunction Verification F15 Series Fetal & Maternal Monitor Service Manual If any one of the results exceeds the reference range, the wireless charging board defection is confirmed. Replacement of the board is recommended. - 47 -...
  • Page 59: Chapter 9 Disassembling The Monitor

    F15 Series Fetal & Maternal Monitor Service Manual Disassembling the Monitor Chapter 9 Disassembling the Monitor WARNING 1 Only qualified service personnel should open the monitor case. 2 Switch off the monitor and remove the AC power cable before disassembling the monitor.
  • Page 60 F15 Series Fetal & Maternal Monitor Service Manual Disassembling the Monitor Upper Assembly Lower Assembly Front Case To disassemble the main unit, 1) Fold the LCD display completely flat. 2) Carefully place the monitor upside down on a flat surface covered with cloth or other protecting pad.
  • Page 61: Disassembling The Lower Assembly

    F15 Series Fetal & Maternal Monitor Service Manual Disassembling the Monitor 1 –Upper assembly 2 –Lower assembly 3 –Ground Wire 4 –Touching/Alarm Indicator/Indicator Connection Cable 5 –Display Connection Wire 6 –Charging Stand Connection Wire (wireless model only) 6 –Switch Connection Wire 7) Disconnect the cables/wires to separate the two parts.
  • Page 62: Replacing The Fan

    F15 Series Fetal & Maternal Monitor Service Manual Disassembling the Monitor Replace the DC-DC transfer board in a reverse procedure. Refer to the following illustration when connecting wires. Ensure they are all correctly connected. 1: Power Output Cable Socket 2: Battery Charging Cable Socket 3:...
  • Page 63: Replacing The Power Module

    F15 Series Fetal & Maternal Monitor Service Manual Disassembling the Monitor 9.4.3 Replacing the Power Module To remove the power module, 1) Separate the upper assembly and the lower assembly. 2) Disconnect all the wires. 3) Remove the4 screws securing the...
  • Page 64: Replacing The Wifi Module

    F15 Series Fetal & Maternal Monitor Service Manual Disassembling the Monitor 9.4.4 Replacing the WIFI Module To remove the WIFI module, 1) Separate the upper assembly and the lower assembly. 2) Remove the 3 screws on the WIFI module. To replace the WIFI module, Replace the WIFI module in a reverse procedure.
  • Page 65: Replacing The Im X12 Parameter Board (Wired Model Only)

    F15 Series Fetal & Maternal Monitor Service Manual Disassembling the Monitor No. Connect to 1 alarm indicator/indicator board 2 touchscreen 3 screen 4 printer head 5 out-of paper checking board 6 motor 7 paper drawer in place checking switch 8 maternal parameter board...
  • Page 66: Replacing The Loudspeaker

    F15 Series Fetal & Maternal Monitor Service Manual Disassembling the Monitor No. Connect to 1 pump connection wire 2 air valve connection wire 3 main control board 4 SpO2 connection wire 5 TEMP connection wire 6 NA 9.4.7 Replacing the Loudspeaker...
  • Page 67: Replacing The Transducer Socket Board (Wired Model Only)

    F15 Series Fetal & Maternal Monitor Service Manual Disassembling the Monitor 9.4.8 Replacing the Transducer Socket Board (wired model only) To remove the transducer socket board, 1) Separate the upper assembly and the lower assembly. 2) Remove the five screws on the transducer socket board.
  • Page 68: Replacing The Printhead

    F15 Series Fetal & Maternal Monitor Service Manual Disassembling the Monitor Replace the bottom interface board in a reverse procedure. Refer to the following illustration when connecting wires. Ensure they are all correctly connected. No. Connect to bottom interface board...
  • Page 69: Replacing The Ap Module (Wireless Model Only)

    F15 Series Fetal & Maternal Monitor Service Manual Disassembling the Monitor CAUTION The thermal printhead is a fragile part. Handle it with care. 9.4.11 Replacing the AP Module (wireless model only) To remove the AP module, 1) Separate the upper assembly and the lower assembly.
  • Page 70: Disassembling The Upper Assembly

    F15 Series Fetal & Maternal Monitor Service Manual Disassembling the Monitor 2) Remove all the cables on the wireless transducer holder bracket. 3) Remove the rubber gasket on the fan. To replace the wireless transducer holder bracket, Replace the wireless transducer holder bracket in a reverse procedure. Refer to the following illustration when connecting wires.
  • Page 71: Chapter 10 Periodic Tests

    F15 Series Fetal & Maternal Monitor Service Manual Periodic Tests Chapter 10 Periodic Tests 10.1 Safety Tests Safety tests should be performed by qualified personnel every 6 to 12 months, and each time after service. Use a standard safety tester to perform safety tests. The tester must comply with IEC/EN 60601-1 or the local standards and statutes.
  • Page 72: Toco Performance Test

    F15 Series Fetal & Maternal Monitor Service Manual Periodic Tests 10.2.2 TOCO Performance Test This test checks the performance of TOCO measurement. 1) Connect the TOCO transducer to the fetal monitor, adjust the pressure test tool to balance and lay the transducer horizontally on the tool surface so that the pressure spot of the tool is right in the middle of the transducer.
  • Page 73: Spo2 Performance Test

    F15 Series Fetal & Maternal Monitor Service Manual Periodic Tests 10.2.4 SpO2 Performance Test This test checks the performance of SpO measurement. Required Equipment: SpO Simulator Connect the equipment and the monitor as shown below: Simulator Sensor Switch on the monitor.
  • Page 74: Fecg Performance Test(Wireless Model Only

    F15 Series Fetal & Maternal Monitor Service Manual Periodic Tests Resistance Box Switch on the monitor. Adjust the resistance value to 32657Ω, 6017Ω and 3602.4Ω in turn. The TEMP numeric value displayed on the screen correspond to 0 º C, 37 º C and 50 º C.
  • Page 75 F15 Series Fetal & Maternal Monitor Service Manual Periodic Tests accepted. - 64 -...
  • Page 76: Appendix 1 Renewal Parts

    F15 Series Fetal & Maternal Monitor Service Manual Renewal Parts Appendix 1 Renewal Parts CAUTION Only connect the renewal parts supplied by EDAN to the monitor. Repair Part Part Number F15 bottom interface board (not 02.03.221006 configured with RF module) F15 bottom interface board 02.03.220833...
  • Page 77: Appendix 2 Connecting To Ob Tracevue System

    F15 Series Fetal & Maternal Monitor Service Manual Connecting to OB TraceVue System Appendix 2 Connecting to OB TraceVue System Preparation Prior to connecting F15/F15Air to the OB TraceVue obstetrical documentation and surveillance system of Philips, make sure that: 1) Net Version of F15/F15Air on the Network Setup interface has this option: PHILIPS.
  • Page 78 F15 Series Fetal & Maternal Monitor Service Manual Connecting to OB TraceVue System Check other bedside monitors in the OB TraceVue system to see if protocol converters (as shown in the following illustrations) are used. Figure A2-3 Protocol Converter Figure A2-4 A Protocol Converter Used on a Monitor If protocol converters are used on other bedside monitors, connect F15/F15Air with the Serial Port to RJ45 Cable #2 (01.13.036124).
  • Page 79 F15 Series Fetal & Maternal Monitor Service Manual Connecting to OB TraceVue System Figure A2-5 Connecting DB9 and RJ45 interfaces of the OB TraceVue system Setting Set the Net Version of F15/F15Air to PHILIPS. (Refer to 5.8 Changing Net Version) Restart F15/F15Air.

Table of Contents